Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 

How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 




 
MRI Forum
'Artifact'
SEARCH FORUM FOR   
 
Result: Searchterm 'Artifact' found in 33 messages
Result Pages: 1 2 3 4 5 6 [7] 
More Results: Database  (166)  News Service  (6)  Resources  (10)  
Forum Overview
 bottom
Brad Johnson

Wed. 26 Jan.05,
03:57

[Reply (1 of 2) to:
'3T Pulsoxymeter'
started by: 'Lynn Dunkan'
on Wed. 3 Nov.04]


 
  Category: 
Equipment

 
3T Pulsoxymeter
Hi Lynn,
we use the pulse oximeter Nonin 8600 for our 1.5 T scanner as well as for the 3T system. I'm not sure whether it is tested for 3 T. At the 3 T system, we run the cable through filter boxes in the shielded scanner room, so there is no safety problem and we do not have artifacts.
It works well also for children.
 View the whole thread
Reader Mail

Tue. 14 Dec.04,
17:25

[Start of:
'MR image of an eye'
1 Reply]


 
  Category: 
Applications and Examinations

 
MR image of an eye
I have a set of MR image of an eye which is acquired from a commercial surface coil. The questions are:

(1) How do I know there is any eye movement during imaging?

(2) How do I remove the artifact if there is any eye movement? Is there a program to do so?

It is deeply appreciated to have help from you.
 View the whole threadReply to this thread
(login or register first)
Schmidt Marga

Sat. 17 Apr.04,
20:41

[Reply (1 of 2) to:
'Moire fringing'
started by: 'Kelly Wimpey'
on Tue. 6 Apr.04]


 
  Category: 
Artifacts

 
Moire fringing
Moire fringes are typical for gradient echo sequences, especially balanced sequences like for example true fisp, bFFE. The reasons are interferences and field inhomogeneities. High order shimming (volume shimming), a more suitable coil, a smaller bandwidth, and scanning not in the edges of the magnetic field will avoid or reduce this artifact. Also parallel imaging techniques can have a positive influence. If this artifact appears, also by using other sequences, a new field adjustment can be necessary (service).
Hope this will help.
 View the whole thread

Result Pages : 1 2 3 4 5 6 [7] 
 top
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Look
      Ups






MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]