  Category: 
Artifacts

 
SPAIR movement artefact
Dear all,
When scanning wrists in our Avanto MR always the penultimate slice in the transversal series shows a movement artefact. This happens with all patients and always the penultimate slice. We found out that using regular fat saturation this does not occur, with SPAIR (which is preferable because of the frequent susceptibility artefacts in arms/wrists) however it happens all the time. The simple solution for our radiology staff is to use normal fat saturation instead of SPAIR, however we are intriged by this problem and were hoping that anyone also experienced this problem and maybe found an explanation.
Because it happens in only 1 slice we are suspecting some kind of interference of the slice selection pulse with the SPAIR pre-pulses.
Any input is appreciated,
thanks in advance, Matthijs
Reply to this thread
(login or register first)
Schmidt Marga

Sat. 17 Apr.04,
20:41

[Reply (1 of 2) to:
'Moire fringing'
started by: 'Kelly Wimpey'
on Tue. 6 Apr.04]


 
  Category: 
Artifacts

 
Moire fringing
Moire fringes are typical for gradient echo sequences, especially balanced sequences like for example true fisp, bFFE. The reasons are interferences and field inhomogeneities. High order shimming (volume shimming), a more suitable coil, a smaller bandwidth, and scanning not in the edges of the magnetic field will avoid or reduce this artifact. Also parallel imaging techniques can have a positive influence. If this artifact appears, also by using other sequences, a new field adjustment can be necessary (service).
Hope this will help.
 View the whole thread

Result Pages : [1] 
 top
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Look
      Ups






MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]