Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 

How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 




 
MRI Forum
'Spin Echo'
SEARCH FORUM FOR   
 
Result: Searchterm 'Spin Echo' found in 12 messages
Result Pages: [1] 2 3 
More Results: Database  (96)  News Service  (1)  
Forum Overview
 bottom
Steven Ford

Fri. 26 Oct.18,
02:29

[Reply (1 of 2) to:
'When to shim?'
started by: 'Matt H'
on Fri. 27 Oct.17]


 
  Category: 
General

 
When to shim?
you ask a complex question. Any sequence that is not a standard spin echo can benefit from shimming, but its not always worth it.

If you are scanning a knee in the center of the magnet, or nearly so, you might not see much difference in gradient echo scans, which in general are quite sensitive to shim problems. But if you are doing fat-water separation or fatsat images, you will see a difference even in this example.

Because hardware varies, its hard to say in one blanket statement for every case. For sure, do it on fatsat or fat-water imaging. The next most sensitive studies are off-center imaging such as shoulders. Beyond that, you should experiment a little. Of course, erring on the side of caution is a good idea.
 
 

Steven Ford
Professional Imaging Services, Inc.
 View the whole thread
Angie Pope

Tue. 23 Oct.18,
16:55

[Start of:
'DWI of the prostate'
0 Reply]


 
  Category: 
Applications and Examinations

 
DWI of the prostate
In need of a protocol for the prostate. Our system is a 1.5 GE Signa Excite 12.0 software.
Under the diffusion options screen the only B-value options are 1000,1500, and 2000.
Should Dual Spin Echo option be selected ?
Main question relates to post processing of the DWI to ADC. Under Functool, what is a good threshold # to choose ?
Reply to this thread
(login or register first)
aaron yonts`

Thu. 19 Sep.13,
07:56

[Reply (1 of 3) to:
'why there is T2* but not T1*'
started by: 'Isa Toni Toni'
on Wed. 24 Jul.13]


 
  Category: 
Sequences and Imaging Parameters

 
why there is T2* but not T1*
i may just be giving general info but a t2* is aquired from a gradient echo compared to a t2 with a spin echo. according to mri in practice a t2* can help show some things a normal t2 would not, such as ms plaques in a cervical spine
 View the whole thread
George LoGuirato

Wed. 7 Apr.10,
13:28

[Reply (1 of 2) to:
'Proton Density Imaging'
started by: 'Jacquelyn Chiffons'
on Thu. 18 Mar.10]


 
  Category: 
Sequences and Imaging Parameters

 
Proton Density Imaging
A Proton Density - like a T1 or T2 - is not considered anything; it is what you make it. Most all PD's are run as fast (turbo) spin echo, but many sites run a true spin echo PD on knees to evaluate the meniscus. (It's a long sequence.)
 View the whole thread
Jacquelyn Chiffons

Thu. 18 Mar.10,
14:51

[Start of:
'Proton Density Imaging'
1 Reply]


 
  Category: 
Sequences and Imaging Parameters

 
Proton Density Imaging
Is Proton Density Imaging considered Fast Spin Echo?
 View the whole threadReply to this thread
(login or register first)

Result Pages : [1] 2 3 
 top
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Look
      Ups






MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 3 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]