Susan Angelico
Thu. 15 Sep.05, 20:07
[Reply (1 of 3) to: '3T imaging parameters' started by: 'Nancy Dudek' on Fri. 26 Aug.05]
Category:
Sequences and Imaging Parameters |
3T imaging parameters |
At 3T the T1 time is longer (so the TR should also be longer) and the T2 time is shorter (TE can be shorter). However, when we got a 3T machine we tested different TR and TE's and found out that in practice the useful range don't has a big difference from that of a 1.5T MR system.
The contrast looks a bit different - that's normal.
In addition, ask the manufacturer.
There are a lot of other parameters that affect the contrast in 3T.
Hope this will help
|
|
View the whole thread | |