Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 

Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 




 
MRI Forum
'Signa'
SEARCH FORUM FOR   
 
Result: Searchterm 'Signa' found in 78 messages
Result Pages: 1 [2] 3 4 5 6 7 8 9 10 11 12 13 14 15 16 
More Results: Database  (382)  News Service  (60)  Resources  (16)  
Forum Overview
 bottom
Kimberly Wilridge

Fri. 26 Jul.13,
16:42

[Start of:
'GE Signa - 17,18,19 Phantoms Found looking for 1'
1 Reply]


 
  Category: 
Coils

 
GE Signa - 17,18,19 Phantoms Found looking for 1
Getting this message on specific coils (CTL) 17,18,19 phantoms found looking for 1. Message never seen before and not sure if coils issue
 View the whole threadReply to this thread
(login or register first)
Giorgos Menikou

Sat. 27 May.06,
07:16

[Start of:
'GE Signa Excite 1.5 T HD spectroscopy fine shimming'
0 Reply]


 
  Category: 
Spectroscopy

 
GE Signa Excite 1.5 T HD spectroscopy fine shimming
Can anyone please give directions for how to use the SpectroScan Tool in spectroscopic sequences.
It is necessary to optimise the shimming in spectroscopy in order to get well defined sharp peaks in the metabolites spectrum.
Reply to this thread
(login or register first)
Gina Key

Sat. 27 Jan.24,
09:33

[Reply (1 of 2) to:
'IDEAL sequence help'
started by: 'gianluca turcatel'
on Mon. 10 Jan.22]


 
  Category: 
General

 
IDEAL sequence help
To calculate the Proton Density Fat Fraction (PDFF) from MRI images of the liver acquired using LMS IDEAL (Iterative Decomposition of Water and Fat with Echo Asymmetry and Least Squares Estimation) with optimized low flip angles, you'll typically have multiple image sets corresponding to different echoes. The fat and water signals are separated based on their chemical shift differences. Look for image sets labeled as Water, Fat, or PDFF (Proton Density Fat Fraction). Depending on the software or DICOM tags used, the images may be explicitly labeled, or you may need to review the sequence parameters to determine which set corresponds to fat or water.
 View the whole thread
Renate Semrau

Mon. 28 Aug.23,
07:23

[Reply (1 of 2) to:
'Improving MEDIC/MERGE/M-ffe and GRE/FFE image quality'
started by: 'Dan Lopez'
on Sun. 27 Aug.23]


 
  Category: 
Sequences and Imaging Parameters

 
Improving MEDIC/MERGE/M-ffe and GRE/FFE image quality
Hi,
Sequences with multiple echoes are unfortunately sensitive to motion and flow artifacts, with the mFFE the echoes are still cumulated afterwards. The use of the head-neck coil provides better SNR. Important are good flow compensation, rest slab, try out foldover direction l/r vs. a/p. Shimming may help, synergy is sometimes better as clear, don't take too many slices, but maybe the signal loss is also a technical problem. Problems in the lung region are common. In short, we use 2D T2 TSE with continuous flow comp.

 View the whole thread
Dan Lopez

Sun. 27 Aug.23,
00:03

[Start of:
'Improving MEDIC/MERGE/M-ffe and GRE/FFE image quality'
1 Reply]


 
  Category: 
Sequences and Imaging Parameters

 
Improving MEDIC/MERGE/M-ffe and GRE/FFE image quality
I work on an Ingenia Elition Philips scanner. I'm having trouble with Axial C-spine sequences for both M-ffe and FFE. On the M-ffe there is always motion in the cord and the contrast is poor. With the FFE the closer we get to the lungs the worse the quality becomes. There is usually loss of signal and the image doesn't look uniform. I've tried a bunch of different parameter combinations to try giving it more signal and contrast while decreasing the motion in the cord but it never seems to look that good. Any suggestions would be appreciated.
 
 

Mffe & FFE
 View the whole threadReply to this thread
(login or register first)

Result Pages : 1 [2] 3 4 5 6 7 8 9 10 11 12 13 14 15 16 
 top
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Look
      Ups






MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]